Understanding FDA's PCCP Guidance — What's Final, What's Still Draft

Two documents, two timelines. Here's which one applies to you.

Updated November 2026: FDA's PCCP guidance for AI-enabled device software is final. The guidance covering all devices is still in draft.

What is a PCCP?

A PCCP lets a manufacturer plan for future device changes and have them pre-cleared as part of the original submission, rather than filing a new 510(k) or PMA supplement for each one.

  1. Aug 2024General PCCP draft issued (all devices)
  2. Dec 2024AI-device PCCP guidance finalized
  3. FY26–27General guidance targeted for finalization (CDRH priority)
Most searches for "FDA PCCP guidance" run into this exact confusion — here's the short version before the detail.
Status
AI-Enabled Device Software FunctionsFinal — December 2024
General PCCP (All Devices)Draft — August 2024
Scope
AI-Enabled Device Software FunctionsAI/ML-driven software functions
General PCCP (All Devices)Any device needing PMA, 510(k), or De Novo — including non-software devices like sutures, IVDs, and patient monitors
Use now?
AI-Enabled Device Software FunctionsYes
General PCCP (All Devices)Preview only — not yet for implementation

Both pathways sit within FDA’s broader premarket cybersecurity guidance expectations for device submissions.

What a PCCP covers

  • Planned Modifications

    The changes you anticipate making to the device.

  • Methodology

    How those changes get developed, validated, and implemented.

  • Impact Assessment

    How each change affects the device's safety and effectiveness.

What the final guidance added

Implications for manufacturers

PCCPs signal a shift toward more flexible regulatory pathways for devices that need frequent updates. That especially helps manufacturers of connected devices, where cybersecurity updates are critical to patient safety and connectivity makes those updates possible. The trade-off is more transparency and documentation up front, which can be a challenge for teams not used to proactive change management.

Intended-use limits

Notably, the guidance emphasizes that modifications under a PCCP must maintain the device’s intended use. Any changes to the device’s intended use or indications for use typically require additional regulatory review, except in very limited circumstances as enumerated in the draft guidance.

Case Study: Implementing a PCCP with Medcrypt’s Cybersecurity Expertise

Medcrypt recently partnered with an MDM to support the development of a PCCP as part of their regulatory submission. The device is connected and will require frequent updates for clinical functionality, adjustments to an algorithm and secure patching. The device is also interoperable. The PCCP that was developed with this MDM supports a variety of clinical functionality updates that do not affect intended use or indications for use. The change types also included minor tweaks to how the algorithm utilized data input from interoperable components in order to handle a variety of input sources that may change over time. Finally, the PCCP demonstrates the manufacturers’ plan for secure patching in order to handle any software updates that are needed to support clinical functionality or new interoperable component interfacing.

Medcrypt reviewed the change types, the modification protocols associated with each change type, the impact assessments, and the device’s overall cybersecurity design to ensure that the PCCP would provide evidence supporting continuing safety, effectiveness, and security of the device.

As a result, the manufacturer successfully submitted their PCCP, ensuring they could implement future cybersecurity updates without requiring additional regulatory submissions for each modification.

FAQ

What is a PCCP?

A predetermined change control plan is part of your marketing submission. It describes the changes you plan to make, how you'll make and test them, and how you'll assess their impact. Once FDA authorizes it, changes made within the plan don't need a new submission.

Is FDA's PCCP guidance final?

It's final for AI-enabled device software functions. The broader guidance that applies to all devices is still a draft.

Do cybersecurity patches need a PCCP?

Many security-only patches don't need a new submission at all. A PCCP helps when a planned change would otherwise trigger one, such as swapping a cryptographic algorithm in a way that affects performance.

How does a PCCP fit with a 524B postmarket plan?

Section 524B requires a plan to monitor, find and fix vulnerabilities. A PCCP can pre-authorize some of the fixes that plan will produce. See our postmarket cybersecurity page.