Updated Nov 2026: For 510(k) and De Novo eSTAR submissions, eSTAR and FDA's technical screening replaced the RTA review.
For 510(k) and De Novo submissions filed through eSTAR, there is no refuse to accept (RTA) review anymore. eSTAR makes sure every required field is filled in, including the cybersecurity section, and FDA's technical screening checks that what's filled in is adequate. PMAs work differently, as the FAQ below explains. This post reflects FDA's current premarket cybersecurity guidance, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions," revised in February 2026.
What replaced RTA for eSTAR submissions
- eSTAR checks completeness. The template won't submit while a required field is blank.
- Technical screening checks adequacy. FDA puts the eSTAR on a technical screening hold if the cybersecurity section lacks accurate responses or relevant attachments.
- eSTAR is required. For 510(k)s since October 1, 2023, and for De Novos since October 1, 2025.
- The cyber device RTA guidance has expired. FDA's cybersecurity FAQ says the RTA policy guidance for cyber devices expired on October 1, 2023.
What triggers a cybersecurity hold
Section 524B applies to cyber devices: devices that include software, can connect to the internet, and could be vulnerable to cybersecurity threats. For these devices, the submission must include:
- A plan to monitor, identify and address postmarket vulnerabilities, including coordinated vulnerability disclosure
- Processes that give reasonable assurance the device and related systems are cybersecure, including postmarket updates and patches
- A software bill of materials covering commercial, open-source and off-the-shelf software components
In eSTAR, inaccurate answers or missing attachments here put the submission on hold at technical screening. For example, a device that lists a USB port but claims no electronic interfaces. FDA's guidance also expects a threat model, a security risk assessment and testing evidence. Gaps there tend to show up as deficiencies later in review.
What happens after a hold
FDA emails you the gaps. You have 180 days to send a complete replacement eSTAR, or the submission is considered withdrawn. Clearing screening doesn't end it: reviewers can still issue cybersecurity deficiencies under Section 524B. What is eSTAR? covers screening in more detail, and Respond to an AINN covers what to do with a deficiency letter.
How to avoid a hold
- Confirm early whether your device is a cyber device under 524B.
- Map your documentation to the guidance before you build the eSTAR. See our clause-by-clause mapping of FDA's premarket cybersecurity guidance.
- Keep your eSTAR answers consistent with your architecture, labeling and SBOM.
- Include each SBOM component's support level and end-of-support date.
- Make sure your threat model, risk assessment and test results trace to each other.
FAQ
What is FDA's refuse-to-accept (RTA) policy for cybersecurity?
RTA is FDA's check that a premarket submission is complete enough to accept for review. For cyber devices, that meant the cybersecurity information Section 524B requires, such as an SBOM, a plan for monitoring and addressing postmarket vulnerabilities, and evidence of secure design. FDA's cybersecurity FAQ says the RTA policy guidance for cyber devices expired on October 1, 2023. For 510(k) and De Novo submissions, which must use eSTAR, there is no RTA review anymore: eSTAR checks that every required field is filled in, and technical screening checks that the answers are adequate.
Where does cybersecurity go in an eSTAR submission?
eSTAR has a dedicated cybersecurity section. FDA's technical screening checks it before substantive review starts, and the eSTAR goes on hold if the section lacks accurate responses or relevant attachments.
How do you avoid a cybersecurity RTA or screening hold?
Check the submission against FDA's premarket cybersecurity guidance before you file. Give accurate answers in the eSTAR cybersecurity section, attach the documents they refer to, and make sure every element 524B requires is easy for a reviewer to find.
Does RTA still apply to PMAs?
Yes, when the PMA isn't filed through eSTAR. eSTAR is voluntary for PMAs, and a PMA filed outside eSTAR still gets FDA's acceptance review, which uses a checklist and can end in a refuse to accept decision, followed by a filing review.
How Medcrypt helps
Our team includes former FDA reviewers. We review your cybersecurity section against the guidance before you file (audit your submission), and we help you respond if FDA puts it on hold. Start with a free one-hour readiness check.